SINGLE
PHASE ISOLATION TRANSFORMERS TMI64 RANGE |
|
SUITABLE FOR USE
IN MEDICAL ENVIRONMENTS |
| FROM
3150 TO 25000VA |
230V
|
230V
|
50/60
Hz |
EN
61558-2-15 |
| CEI
64-4 |
| CEI
96-2 |
| BS
3535 |
Open type
Protection IP00
Isolation class H (Ta max 45°C)
Separate copper windings cl. H
Resin impregnated and tropicalised
Double earthed interwinding copper screen
Central tap on secondary winding
Connections on terminal blocks to 10kVA (or 63A)
Coonections on brass bolts from 12,5kVA (or from 64A)
Encapsulated in metal enclosure on request
|


|

|
| power |
part
no. |
losses |
short
circ. |
regulation |
dimensions
mm. |
DWG |
weight |
box |
| VA |
W |
voltage
% |
% |
A |
B |
C |
D |
E |
F |
kg. |
code |
| 3150 |
TMI6403 |
130 |
3.5 |
3.0 |
240 |
230 |
320 |
200 |
125 |
12 |
A |
34.0 |
SA02 |
| 4000 |
TMI6404 |
145 |
2.9 |
2.6 |
240 |
240 |
320 |
200 |
135 |
12 |
A |
38.0 |
SA02 |
| 5000 |
TMI6405 |
180 |
2.7 |
2.6 |
240 |
250 |
320 |
200 |
145 |
12 |
A |
44.0 |
SA02 |
| 6300 |
TMI6406 |
200 |
2.7 |
2.6 |
280 |
230 |
400 |
220 |
145 |
12 |
A |
49.0 |
SA03 |
| 7500 |
TMI6408 |
210 |
2.7 |
2.6 |
280 |
240 |
400 |
220 |
155 |
12 |
A |
60.0 |
SA03 |
| 10000 |
TMI6410 |
280 |
2.5 |
2.3 |
280 |
250 |
400 |
220 |
165 |
12 |
A |
65.0 |
SA03 |
| 15000 |
TMI6415 |
400 |
2.5 |
2.2 |
320 |
260 |
450 |
260 |
185 |
12 |
B |
92.0 |
SA04 |
| 20000 |
TMI6420 |
600 |
2.4 |
2.0 |
320 |
280 |
450 |
260 |
205 |
12 |
B |
116.0 |
SA04 |
| 25000 |
TMI6425 |
700 |
2.1 |
2.0 |
320 |
280 |
450 |
260 |
205 |
12 |
B |
131.0 |
SA04 |
| ASK
FOR CODES AND SPECIFICATION OF HIGHER RATINGS |
|
THE PART
NO. REPRESENTS THE STANDARD PRODUCT WITH INOUT/OUTPUT VOLTAGE
230V. FOR DIFFERENT VOLTAGES, THE ORDER SHALL INCLUDE THE
REQUIRED INPUT/OUTPUT VOLTAGE |
|
|
A |

|
|
B |

|
|
|
Data shown